8 – Peroxide Fusion - ICPMS/ICP

Samples are fused with sodium peroxide and undergo an acid dissolution. Samples are then analyzed by Perkin Elmer Sciex ELAN 6000, 6100 or 9000 ICP/MS. A fused blank is run in triplicate for every 22 samples. Fused controls and standards are run after the 22 samples. Fused duplicates are run every 10 samples. The instrument is recalibrated every 44 samples.

Element

Detection Limit

 

Element

Detection Limit

Al

0.001  %

 

Nb

0.001 %

As 0.001-10%
Ni 0.001 %
Be 0.001 %
Pb 0.001 %
Bi 0.001 %
Re 0.001 %
Ca 0.01 %
S 0.01 %
Co 0.001 %
Sb 0.002 %
Cr 0.01 %
Se 0.001 %
Cs 0.001 %
Si 0.01 - 47 %
Cu 0.001 %
Sn 0.001 %
Fe 0.05 %
Ta 0.001 %
Ga 0.001 %
Te 0.001 %
Ge 0.001 %
Th 0.001 %
In 0.001 %
Ti 0.01 %
K 0.1 %
Tl 0.001 %
Li 0.001 %
U 0.001 %
Mg 0.01 %
W 0.001 %
Mn 0.01 %
Zn 0.001 %
Mo 0.001  %


     

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 