Weblims     OWA     Home     Contact Us    

Actlabs
Back to Home


29 of 150    Prev    Next     

1F - 4-Acid "Near Total" Digestion - ICP

A 0.25 g sample is digested with four acids beginning with hydrofluoric, followed by a mixture of nitric and perchloric acids. This is then heated using precise programmer controlled heating in several ramping and holding cycles which takes the samples to incipient dryness. After incipient dryness is attained, samples are brought back into solution using aqua regia.


With this digestion, certain phases may be only partially solubilized. These phases include zircon, monazite, sphene, gahnite, chromite, cassiterite, rutile and barite. Ag greater than 100 ppm and Pb greater than 5000 ppm should be assayed as high levels may not be solubilized. Only sulphide sulfur will be solubilized.


The samples are then analyzed using a Varian ICP.  QC for the digestion is 14% for each batch, 5 method reagent blanks, 10 in-house controls, 10 samples duplicates, and 8 certified reference materials. An additional 13% QC is performed as part of the instrumental analysis to ensure quality in the areas of instrumental drift. 

 

Code 1F - Elements and Detection Limits (ppm)

Element

Detection Limit

Upper Limit

Element

Detection Limit

Upper Limit

Element

Detection Limit

Upper Limit

Ag

0.3

100

Fe

0.01%

50%

Pb

3

5,000

Al

0.01%

50

K

0.01%

10%

S

0.01%

20%

Be

1

10,000

Mg

0.01%

50%

Sr

1

10,000

Bi

2

10,000

Mn

1

100,000

Ti

0.01%

10%

Ca

0.01%

70%

Mo

1

10,000

V

2

10,000

Cd

0.3

2,000

Na

0.01%

10%

Y

2

-

Co

1

10,000

Ni

1

10,000

Zn

1

10,000

Cu

1

10,000

P

0.001%

10%

 

Note:  

Extraction of each element by 4-Acid Digestion is dependent on mineralogy.

 


Activation Laboratories Ltd. | 41 Bittern Street, Ancaster, Ontario, L9G 4V5, Canada | TF: +1.888.228.5227