4B2 - STD - Lithium Metaborate/Tetraborate Fusion - ICP/MS

Samples fused under code 4B2 are diluted and analyzed by Perkin Elmer Sciex ELAN 6000, 6100 or 9000 ICP/MS. Three blanks and five controls (three before the sample group and two after) are analyzed per group of samples. Duplicates are fused and analyzed every 15 samples. Instrument is recalibrated every 40 samples.

Code 4B2-STD  Elements and Detection Limits (ppm)

Element

Detection Limit

Upper Limit

Element

Detection Limit

Upper Limit

Element

Detection Limit

Upper Limit

Ag

0.5

100

Hf

0.2

1,000

Sn

1

1,000

As

5

2,000

Ho

0.1

1,000

Sr

2

10,000

Ba

3

300,000

In

0.2

200

Ta

0.1

500

Bi

0.4

2,000

La

0.1

2,000

Tb

0.1

1,000

Ce

0.1

3,000

Lu

0.04

1,000

Th

0.1

2,000

Co

1

1,000

Mo

2

100

Tl

0.1

1,000

Cr

20

10,000

Nb

1

1,000

Tm

0.05

1,000

Cs

0.5

1,000

Nd

0.1

2,000

U

0.1

1,000

Cu

10

10,000

Ni

20

10,000

V

5

5,000

Dy

0.1

1,000

Pb

5

10,000

W

1

5,000

Er

0.1

1,000

Pr

0.5

1,000

Y

1

1,000

Eu

0.05

1,000

Rb

2

1,000

Yb

0.1

1,000

Ga

1

500

Sb

0.5

200

Zn

30

10,000

Gd

0.1

1,000

Sm

0.1

1,000

Zr

5

10,000

Ge

1

500

 

 

Typical ICP/MS Standards Analysis (119 measurements)

Element

W2

Cert.

 

Element

W2

Cert.

 

Element

W2

Cert.

 

Element

W2

Cert.

V

256

262

 

Y

21

24

 

Ce

24

24

 

Yb

2.06

20.5

Cr

90

93

 

Zr

99

94

 

Pr

2.5

5.9?

 

Lu

0.33

0.33

Co

44

44

 

Nb

7.5

7.9

 

Nd

14

14

 

Hf

2.64

2.56

Ni

67

70

 

Mo

0.7

0.6

 

Sm

3.38

3.25

 

Ta

0.5

0.5

Cu

105

103

 

Ag

<0.5

0.05

 

Eu

1.1

1.1

 

W

<0.2

0.3

Zn

72

77

 

In

<0.2

-

 

Gd

3.5

3.6

 

Tl

0.1

0.2

Ga

18

20

 

Sn

<0.5

-

 

Tb

0.62

0.63

 

Pb

8

9.3

Ge

2

1

 

Sb

0.78

0.79

 

Dy

3.8

3.8

 

Bi

<0.05

0.03

As

<5

1.24

 

Cs

0.95

0.99

 

Ho

0.76

0.76

 

Th

2.3

2.5

Rb

20

20

 

Ba

164

182

 

Er

2.3

2.5

 

U

0.49

0.53

Sr

193

194

 

La

11.3

11.4

 

Tm

0.32

0.38

 

 

 

 

 

Code 4B2-STD options:

Code 4B2-STDQuant:  Although intended primarily for unmineralized samples, mineralized samples can be analyzed. However data may be semiquantitative for chalcophile elements (Ag, As, Bi, Co, Cu, Mo, Ni, Pb, Sb, Sn, W and Zn). 

A 1 g sample is digested with aqua regia and diluted to 250 ml volumetrically. Appropriate international reference materials for the metals of interest are digested at the same time. The samples and standards are analyzed on a Varian Vista 735 or a Thermo ICAP 6500 ICP.

Code 4B1 Recommended for accurate levels of base metals (Cu, Pb, Zn, Ni and Ag).

Code 4B-INAA is recommended for As, Bs, high W >100 ppm and Cr > 1,000 ppm. 

Code 5D : recommended for Sn >50 ppm.


Activation Laboratories Ltd. | 41 Bittern Street, Ancaster, Ontario, L9G 4V5, Canada | TF: +1.888.228.5227