6B - Ion Chromatography

Un-acidified water samples are analyzed using the DIONEX DX-120 Ion Chromatography System to determine and quantify a group of seven anions. This analysis is applicable to concentrations less than 75 mg/L for Cl, NO2 and NO3; less than 50 mg/L for F; less than 125 mg/L for PO4; less than 250 mg/L for Br and less than 375 mg/L for SO4. Samples exceeding this range (with high total dissolved solids) must be diluted to avoid over-saturation. Measurement uncertainty is evaluated and controlled by an appropriate quality assurance program, including the use of regular laboratory duplicates of samples and verification of the precision/calibration of the instrument through regular runs of various primary dilution standard solutions

 

Code 6B - Ion Chromatography

Anion

Detection

Limit (mg/L)

Fluoride (F)

0.01

Chloride (Cl)

0.03

Bromide (Br)

0.03

Nitrite (NO2 )

0.01

Nitrate (NO3 )

0.01

Phosphate (PO4 )

0.02

Sulphate (SO4 )

0.03

 

Reference:
EPA Method 300.0

 


Activation Laboratories Ltd. | 41 Bittern Street, Ancaster, Ontario, L9G 4V5, Canada | TF: +1.888.228.5227