4F - H Total

A 0.3 g sample is thermally decomposed in a resistance furnace in a pure nitrogen environment at 1000 °C, using an ELTRA CW-800, directly releasing H2O, which includes both H2O- and H2O+. H2O absorbs IR energy at a precise wavelength within the IR spectrum. Energy from the IR source is absorbed as the gas passes through the cell, preventing it from reaching the IR detector. All other IR energy is prevented from reaching the IR detector by a narrow bandpass filter. Because of the filter, the absorption of IR energy can be attributed only to water (H2O). The concentration of H2O is detected as a reduction in the level of energy at the detector. Total H is calculated from H2O.

 

Analysis

Method

Detection Limit

Total H

Infrared, Calc.

0.01%